BC-10 Coil Protocols
GE Healthcare BC-10 MRI Coil – 3.0T Protocols
Imaging Parameters |
3T – T1 Weighted |
3T Intermediate/T2 Weighted |
Plane |
Oblique |
Oblique |
Mode |
2D |
2D |
Pulse Sequence |
SE |
FSE-XL |
Imaging Options |
EDR |
TRF, Fast, ZIP512 |
SAT |
As Needed |
As Needed |
Scan Timing |
3T – T1 Weighted |
3T Intermediate/T2 Weighted |
Echoes |
1 |
1 |
TE |
Min Full |
45.0 |
TR |
800 |
3350.0 |
ETL |
|
12 |
Bandwidth (kHz) |
±20.73 |
±31.25 |
Additional Parameters |
3T – T1 Weighted |
3T Intermediate/T2 Weighted |
Frequency Encoding samples |
384 |
256 |
Phase Encoding samples |
256 |
256 |
Matrix/NEX |
384 x 256/2 |
256 x 256/2 |
Phase FOV |
0.75 |
0.75 |
Freq DIR |
R/L |
R/L |
Phase Correct |
No |
Yes |
Scanning Range |
3T – T1 Weighted |
3T Intermediate/T2 Weighted |
FOV (cm) |
10 |
10 |
Slice Thickness (mm) |
2 |
2 |
Spacing (mm) |
1 |
1 |
GE Healthcare BC-10 MRI Coil 1.5T Protocol
Imaging Parameters |
1.5T – T1 Weighted |
1.5T Intermediate/ T2 Weighted |
Plane |
Oblique |
Oblique |
Mode |
2D |
2D |
Pulse Sequence |
SE |
FSE-XL |
Imaging Options |
None |
TRF, Fast, ZIP512 |
SAT |
As Needed |
As Needed |
Scan Timing |
1.5T – T1 Weighted |
1.5T Intermediate/ T2 Weighted |
Echoes |
1 |
1 |
TE |
Min Full |
45.0 |
TR |
500-600 |
3000.0 |
ETL |
|
10 |
Bandwidth (kHz) |
±16 |
±31.25 |
Additional Parameters |
1.5T – T1 Weighted |
1.5T Intermediate/ T2 Weighted |
Frequency Encoding samples |
384 |
256 |
Phase Encoding samples |
256 |
256 |
Matrix/NEX |
384 x 256/ 2 |
256 x 256/ 2 |
Phase FOV |
0.75 |
0.75 |
Freq DIR |
R/L |
R/L |
Phase Correct |
No |
Yes |
Scanning Range |
1.5T – T1 Weighted |
1.5T Intermediate/ T2 Weighted |
FOV (cm) |
10 |
10 |
Slice Thickness (mm) |
3 |
3 |
Spacing (mm) |
0 |
0 |