GE Healthcare BC-10 MRI Coil – 3.0T Protocols

 

Imaging Parameters 3T – T1 Weighted 3T Intermediate/T2 Weighted
Plane Oblique Oblique
Mode 2D 2D
Pulse Sequence SE FSE-XL
Imaging Options EDR TRF, Fast, ZIP512
SAT As Needed As Needed

 

Scan Timing 3T – T1 Weighted 3T Intermediate/T2 Weighted
Echoes 1 1
TE Min Full 45.0
TR 800 3350.0
ETL 12
Bandwidth (kHz) ±20.73 ±31.25

 

Additional Parameters 3T – T1 Weighted 3T Intermediate/T2 Weighted
Frequency Encoding samples 384 256
Phase Encoding samples 256 256
Matrix/NEX 384 x 256/2 256 x 256/2
Phase FOV 0.75 0.75
Freq DIR R/L R/L
Phase Correct No Yes

 

Scanning Range 3T – T1 Weighted 3T Intermediate/T2 Weighted
FOV (cm) 10 10
Slice Thickness (mm) 2 2
Spacing (mm) 1 1

 

GE Healthcare BC-10 MRI Coil 1.5T Protocol

Imaging Parameters 1.5T – T1 Weighted 1.5T Intermediate/ T2 Weighted
Plane Oblique Oblique
Mode 2D 2D
Pulse Sequence SE FSE-XL
Imaging Options None TRF, Fast, ZIP512
SAT As Needed As Needed

 

Scan Timing 1.5T – T1 Weighted 1.5T Intermediate/ T2 Weighted
Echoes 1 1
TE Min Full 45.0
TR 500-600 3000.0
ETL 10
Bandwidth (kHz) ±16 ±31.25

 

Additional Parameters 1.5T – T1 Weighted 1.5T Intermediate/ T2 Weighted
Frequency Encoding samples 384 256
Phase Encoding samples 256 256
Matrix/NEX 384 x 256/ 2 256 x 256/ 2
Phase FOV 0.75 0.75
Freq DIR R/L R/L
Phase Correct No Yes

 

Scanning Range 1.5T – T1 Weighted 1.5T Intermediate/ T2 Weighted
FOV (cm) 10 10
Slice Thickness (mm) 3 3
Spacing (mm) 0 0